Title :
Development of blood vessel viewer using multi spectral imaging technology
Author :
Nagaoka, Takashi ; Nakamura, Atsushi ; Aizawa, Katsuo ; Ohtsubo, Shinya ; Mochizuki, Toru ; Kiyohara, Yoshio ; Sota, Takayuki
Author_Institution :
Shizuoka Cancer Center Research Institute, Cancer Diagnostic Research Division, 411-8777, JAPAN
Abstract :
Using multi spectral imaging technology, which enable us to simultaneously obtain spatial and wavelength information, we have newly developed blood vessel viewer. Our device is composed of the light source, the lens or the endoscope, the optical device, CCD camera, and PC. As an evaluation of our device, we measured the skin diffuse reflectance spectrum of the human hand. The blood vessel observed by our device was corresponding to an anatomy finding. However, even if Principal Component Analysis was used, the reflection of horny was not able to be removed completely.
Keywords :
Biomedical imaging; Blood vessels; Charge coupled devices; Charge-coupled image sensors; Endoscopes; Lenses; Light sources; Optical devices; Optical imaging; Wavelength measurement; Hand; Humans; Image Processing, Computer-Assisted; Optical Devices; Principal Component Analysis; Skin; Spectrum Analysis;
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location :
Vancouver, BC
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2008.4649386