DocumentCode :
3067316
Title :
Nonlinear analysis of the ECG during atrial fibrillation in patients for low energy internal cardioversion
Author :
Diaz, J. ; Gonzalez, C. ; Escalona, O. ; Glover, B.M. ; Manoharan, G.
Author_Institution :
Universidad Simón BolÃ\xadvar, Venezuela
fYear :
2008
fDate :
20-25 Aug. 2008
Firstpage :
1619
Lastpage :
1622
Abstract :
The goal of this study was to investigate the usefulness of nonlinear analysis in determining the success of low energy internal cardioversion (IC) in patients with atrial fibrillation (AF). Nonlinear analysis has previously been used for characterizing AF patterns, and spontaneous termination in its paroxysmal form. However, the relationship between the probability to restore sinus rhythm by IC and quantitative nonlinear analysis based electrocardiographic (ECG) markers has not been explored before. Thirty nine patients with AF, for elective DC cardioversion at the Royal Victoria Hospital in Belfast, were included in this study. One catheter was positioned in the right atrial appendage and another in the coronary sinus, to deliver a biphasic shock waveform. A voltage step-up protocol (50–300 V) was used for patient cardioversion. Residual atrial fibrillatory signal (RAFS) was derived from 60 seconds of surface ECG from defibrillator pads, prior to shock delivery, by bandpass filtering and ventricular activity (QRST) cancellation. QRST complexes were cancelled using a recursive least squared (RLS) adaptive filter. The maximal Lyapunov exponent (λ), correlation dimension (course grained estimation, CDcg) and approximate entropy (ApEn) were extracted from the RAFS. These variables were calculated from 10 s of the RAFS before shock delivery. 26 patients were successfully cardioverted, employing a maximum energy of 11.84 joules. A lower λ (0.037±0.006 vs. 0.044±0.008, P=0.01) and CDcg (5.552±2.075 vs. 6.592±1.130, P=0.049) were found in successfully cardioverted patients than in those non successful ones, with an energy ≤3 joules. Also, there was a significant positive correlation between the minimal defibrillation energy and λ (r=0.483, P=0.013) in cardioverted patients. In conclusion, complexity analysis of the RAFS is useful for assessing the prospective efficacy of internal low energy cardiovers- ion of patients with atrial fibrillation.
Keywords :
Atrial fibrillation; Cardiology; Catheters; Electric shock; Electrocardiography; Hospitals; Pattern analysis; Protocols; Rhythm; Voltage; Atrial Fibrillation; Biomedical Engineering; Defibrillators, Implantable; Electric Countershock; Electrocardiography; Humans; Models, Cardiovascular; Nonlinear Dynamics; Signal Processing, Computer-Assisted;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location :
Vancouver, BC
ISSN :
1557-170X
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2008.4649483