Title :
A multi-channel instrumentation system for biosignal recording
Author :
Yu, Hong ; Li, Pengfei ; Xiao, Zhiming ; Peng, Chung-Ching ; Bashirullah, Rizwan
Author_Institution :
Department of Electrical and Computer Engineering, University of Florida, Gainesville, 32611, USA
Abstract :
This paper reports a highly integrated battery operated multi-channel instrumentation system intended for physiological signal recording. The mixed signal IC has been fabricated in standard 0.5μm 5V 3M-2P CMOS process and features 32 instrumentation amplifiers, four 8b SAR ADCs, a wireless power interface with Li-ion battery charger, low power bidirectional telemetry and FSM controller with power gating control for improved energy efficiency. The chip measures 3.2mm by 4.8mm and dissipates ∼2.1mW when fully operational.
Keywords :
Battery charge measurement; Battery management systems; CMOS integrated circuits; CMOS process; Digital signal processing chips; Electrodes; Energy management; Instruments; Semiconductor device measurement; Signal processing; Amplifiers, Electronic; Analog-Digital Conversion; Electric Power Supplies; Electrophysiology; Equipment Design; Signal Processing, Computer-Assisted; Telemetry;
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location :
Vancouver, BC
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2008.4649587