• DocumentCode
    3070620
  • Title

    Development of a simple stance-control system for persons with poliomyelitis and an associated gait analysis

  • Author

    Nakanishi, Yoshie ; Kato, Noriaki ; Wada, Futoshi ; Hachisuka, Kenji ; Watanabe, Yasuhiro ; Okumura, Katsuhiro ; Arai, Mitsuo

  • Author_Institution
    Dept. of Rehabilitation Med., Univ. of Occupational & Environ. Health, Kitakyushu, Japan
  • fYear
    2012
  • fDate
    1-4 July 2012
  • Firstpage
    169
  • Lastpage
    173
  • Abstract
    A simple stance-control system (SSCS) was developed, which is a new device attached to the upper lateral frame of a carbon fiber-reinforced plastic knee-ankle-foot orthosis (carbon KAFO). The SSCS consisted of two pressure sensors beneath the heel and metatarsal phalangeal joints, connecting cables, microcomputer, motor controlling the lever of Swiss-lock knee joint, and battery box. Normally the SSCS is off, and the Swiss-lock knee joint is mechanically fixed in the extended position, whereas the SSCS is on during the swing phase according to the information from the pressure sensors, and the Swiss-lock knee joint is released by pulling up the lever. Therefore, a patient using a carbon KAFO with SSCS can flex his/her knee joint during the swing phase and to maintain it in the extended position during the stance phase. The safety while using the SSCS was confirmed in four healthy men and the gait pattern was measured by using a three-dimensional motion analysis and force plate system. These results revealed that the SSCS attached to a carbon KAFO did not unlock the Swiss-lock knee joint on the five conditions in daily living activities, and enabled a patient to control his/her lower limb during the stance phase and flex it during the swing phase.
  • Keywords
    biomedical equipment; biomedical materials; bone; carbon fibre reinforced plastics; diseases; gait analysis; medical control systems; microcomputers; orthopaedics; orthotics; pressure sensors; prosthetics; Swiss-lock knee joint; battery box; carbon fiber-reinforced plastic knee-ankle-foot orthosis; connecting cables; force plate system; gait analysis; gait pattern; lower limb; metatarsal phalangeal joints; microcomputer; poliomyelitis; pressure sensors; simple stance-control system; stance phase; swing phase; three-dimensional motion analysis; upper lateral frame; Biomechanics; Energy measurement; Flexible printed circuits; Footwear; Joints; Knee; Legged locomotion; a carbon fiber-reinforced plastic knee-ankle-foot orthosis; a simple stance-control system; kinematic gait analysis; safety performance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Complex Medical Engineering (CME), 2012 ICME International Conference on
  • Conference_Location
    Kobe
  • Print_ISBN
    978-1-4673-1617-0
  • Type

    conf

  • DOI
    10.1109/ICCME.2012.6275704
  • Filename
    6275704