DocumentCode
3074383
Title
Flexible polymer substrate and tungsten microelectrode array for an implantable neural recording system
Author
Patrick, Erin ; Sankar, Viswanath ; Rowe, William ; Yen, Sheng-Feng ; Sanchez, Justin C. ; Nishida, Toshikazu
Author_Institution
Department of Electrical and Computer Engineering, University of Florida, Gainesville, 32611 USA
fYear
2008
fDate
20-25 Aug. 2008
Firstpage
3158
Lastpage
3161
Abstract
This paper describes the process flow and testing of a substrate for a fully implantable neural recording system. Tungsten microwires are hybrid-packaged on a micromachined flexible polymer substrate forming an intracortical microelectrode array for brain machine interfaces. The microelectrode array is characterized on the bench top and tested in vivo. The microelectrode noise floor is less than 2 μV and acute recording results show a signal to noise ratio of 9.9–17.3 dB. The technique of hybrid fabrication of the electrodes on a flexible substrate provides a general platform for the development of an implantable neural recording system
Keywords
Capacitive sensors; Electrodes; In vivo; Microelectrodes; Neurons; Polymer films; Signal to noise ratio; Skull; Substrates; Tungsten; Algorithms; Animals; Biocompatible Materials; Electrodes; Electrodes, Implanted; Electrophysiology; Equipment Design; Humans; Male; Miniaturization; Models, Statistical; Polymers; Rats; Rats, Sprague-Dawley; Tungsten;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location
Vancouver, BC
ISSN
1557-170X
Print_ISBN
978-1-4244-1814-5
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2008.4649874
Filename
4649874
Link To Document