DocumentCode
307477
Title
Internal atrial defibrillation: the importance of lead location, waveform morphology, and shock synchronization
Author
Cooper, Randolph A S
Author_Institution
Cardiac Rhythm Manage. Lab., Alabama Univ., Birmingham, AL, USA
Volume
1
fYear
1995
fDate
20-25 Sep 1995
Firstpage
265
Abstract
Atrial fibrillation (AF) remains the most common arrhythmia encountered in clinical medicine and is associated with significant morbidity and mortality. Electrical cardioversion remains a very effective method for the acute termination of AF. However, the standard method for atrial cardioversion requires shock strengths of 50 to 400 joules delivered across transthoracic cutaneous electrodes. These shocks strengths can be detrimental to the heart as well as painful to the patient. Until recently, there was little available information concerning the technique of internal atrial defibrillation in either animal models or humans. Current research effort have shown that this method usually requires much lower energies, in the 1-5 joule range and requires only sedation and not anesthesia in humans. Over the past 5-6 years the interest in internal atrial defibrillation has greatly grown and now the feasibility of an implantable atrial defibrillator appears soon to be a reality. This manuscript discusses some of the more recent animal and human studies which have looked at the efficacy as well as the safety of internal atrial defibrillation of AF
Keywords
bioelectric phenomena; cardiology; defibrillators; health hazards; safety; 1 to 5 J; 50 to 400 J; animal models; arrhythmia; clinical medicine; electrical cardioversion; heart; humans; implantable atrial defibrillator; internal atrial defibrillation; lead location; morbidity; mortality; safety; sedation; shock strengths; shock synchronization; transthoracic cutaneous electrodes; waveform morphology; Anesthesia; Animals; Atrial fibrillation; Biomedical electrodes; Cardiology; Defibrillation; Electric shock; Heart; Humans; Safety;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location
Montreal, Que.
Print_ISBN
0-7803-2475-7
Type
conf
DOI
10.1109/IEMBS.1995.575102
Filename
575102
Link To Document