Effect of electrode polarization on sensing performance in transvenous defibrillation leads
Author
Hoffmann, Drew A. ; Yang, Weiqun
Author_Institution
Ventritex Inc., Sunnyvale, CA, USA
Volume
1
fYear
1995
fDate
20-25 Sep 1995
Firstpage
309
Abstract
Electrode polarization has been implicated as a possible cause of post-shock electrogram attenuation in transvenous defibrillation electrodes. An in-vitro mapping experiment designed to simulate the electrical activity of the heart is described. Using this technique, electrochemical polarization effects are shown to have a minor influence on post-shock signal amplitude
Keywords
defibrillators; electric potential; electrocardiography; patient monitoring; automatic implantable cardioverter/defibrillator; electrical activity; electrochemical polarization effects; electrode polarization; heart; in-vitro mapping experiment; intracardiac electrogram; post-shock electrogram attenuation; post-shock signal amplitude; sensing performance; transvenous defibrillation leads; Attenuation; Cardiology; Defibrillation; Electric shock; Electrodes; Heart; In vitro; Polarization; Signal generators; Voltage;