Ventricular evoked response measurements from pacing electrodes
Author
Soykan, Orhan ; Schuelke, Robert ; Meador, John ; Morrison, Scott ; Blow, Brian
Author_Institution
Medtronic Inc., Fridley, MN, USA
Volume
1
fYear
1995
fDate
20-25 Sep 1995
Firstpage
331
Abstract
This work presents the results of a study using ventricular electrogram signals to identify the heart´s evoked response following an electrical stimulus from a pacing electrode. Acute data were collected in vivo from seven human subjects and analyzed offline. The ability to recognize the evoked response following the pacing pulse was characterized as a function of the parameters of the pacing pulse. Although the characteristics of the evoked response were dependent on the patient, inter-patient data became consistent when the time derivative of the electrocardiogram was considered
Keywords
electrocardiography; medical signal processing; pacemakers; pattern classification; cardiac pacemakers; electrical stimulus; electrocardiogram; heart; human subjects; inter-patient data; pacing electrodes; pacing pulse; sensing; time derivative; ventricular electrogram signals; ventricular evoked response measurements; Data analysis; Electrocardiography; Electrodes; Heart; Myocardium; Pacemakers; Polarization; Pulse measurements; Signal processing; Time measurement;