Title :
A novel limit alarmer for reliable intensive care monitoring
Author :
Mäkivirta, A. ; Kari, A. ; Koski, E. ; Sukuvaara, T.
Author_Institution :
Med. Eng. Lab., Tech. Res. Centre of Finland, Tampere, Finland
Abstract :
A median filter is combined with narrow adjustable monitoring limits and wide physiologically determined alarm limits. This ensures both quick alarming in the case of a catastrophic failure and reliable therapy monitoring with alarms. Results of the clinical assessment of this dual-limit alarmer are provided. Results of the clinical evaluation indicate a significant reduction in the frequency of false alarm (63% per patient hour)
Keywords :
alarm systems; patient care; patient monitoring; catastrophic failure; false alarm frequency; limit alarmer; median filter; physiologically determined alarm limits; reliable intensive care monitoring; therapy monitoring; Biomedical monitoring; Condition monitoring; Delay effects; Delay estimation; Filtering; Filters; Medical treatment; Patient monitoring; Physiology; Stability;