            Abstract : 
In the application of 3D-volume reconstruction from 2D image slices acquired by a confocal scanning laser microscope errors arise due to refractive index mismatch. This paper presents a brief theoretical analysis of this error that can lead to miscalculating a specimen´s 3D geometry. A comparison between the error analysis and experimental measurements is also provided
         
        
            Keywords : 
biological techniques; biology computing; error analysis; image reconstruction; measurement errors; medical image processing; optical images; optical microscopy; physics computing; refractive index; 2D image slices; 3D confocal scanning laser microscopy; 3D geometry; 3D-volume reconstruction; confocal scanning laser microscope; error analysis; refractive index mismatch; volumetric image distortion;
         
        
        
        