Title
Inaccurate assessment of Doppler-derived mean pressure gradients by commercially available computer programs
Author
Sharif, D. ; Abinader, E.G. ; Goldhammer, E.
Author_Institution
Bnai Zion Med. Center, Haifa, Israel
fYear
1989
fDate
19-22 Sep 1989
Firstpage
375
Abstract
Summary form only given. Computer programs incorporated with Doppler echocardiographic machines and other available programs for offline analysis provide estimation of mean pressure gradients. The accuracy of three programs in the assessment of mean gradients was evaluated. A linear relationship was found between program-derived mean pressure gradients and false gradient calculated as 4×(mean velocity)2 (r =1). Digitalization of the envelope of the Doppler signal and calculation of the mean of the (4×velocity 2) function (the true Doppler-derived pressure gradient) did not correlate with program-calculated false pressure gradients. Errors in mean pressure gradients were sizeable, especially in patients with aortic stenosis. It is concluded that the assumption that mean of squares=square of means may be inaccurate in the assessment of mean pressure gradients and that commercial computer programs should be carefully evaluated before the provided estimate of mean gradients is accepted
Keywords
Doppler effect; biomedical ultrasonics; cardiology; haemodynamics; measurement errors; medical diagnostic computing; pressure measurement; Doppler echocardiographic machines; Doppler-derived mean pressure gradients; aortic stenosis; blood pressure; commercially available computer programs; false pressure gradients; offline analysis; Application software; Blood; Cardiology; Catheterization; Computer errors; Equations; Hemodynamics; Lesions;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology 1989, Proceedings.
Conference_Location
Jerusalem
Print_ISBN
0-8186-2114-1
Type
conf
DOI
10.1109/CIC.1989.130570
Filename
130570
Link To Document