Title :
Bandwidth measurement of the conductance catheter system
Author :
Chen, Chieh-En ; Wei, Chia-Ling
Author_Institution :
Department of Electrical Engineering, National Cheng Kung University, Tainan, Taiwan
fYear :
2008
fDate :
20-25 Aug. 2008
Firstpage :
4859
Lastpage :
4862
Abstract :
Conductance catheter system is used to estimate real-time ventricular volume by measuring the time-varying ventricular conductance/admittance. However, the system is generally calibrated only with known resistors, while neither the frequency response nor the bandwidth of the system is calibrated and measured. The main difficulty of measuring its bandwidth is that the sensed signal of the conductance catheter system, which can be viewed as the input of the system, is a modulated signal, rather than a typical sin wave. Therefore, its bandwidth cannot be measured by typical frequency response analyzers, since they are designed for pure sin-wave input. The waveform of the sensed signal is analyzed. Moreover, a waveform simulator designed to mimic the sensed signal is presented. It can be used to measure the bandwidth of conductance catheter system.
Keywords :
Admittance measurement; Bandwidth; Catheters; Frequency measurement; Frequency response; Real time systems; Resistors; Signal analysis; Time varying systems; Volume measurement; Cardiac Volume; Cardiography, Impedance; Equipment Design; Equipment Failure Analysis; Heart Catheterization; Humans; Reproducibility of Results; Sensitivity and Specificity;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location :
Vancouver, BC
ISSN :
1557-170X
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2008.4650302
Filename :
4650302
Link To Document :
بازگشت