A novel method to estimate the aortic pressure waveform using B-mode ultrasound images acquired from a suprasternal view
Author :
Leung, Mande T. ; Dumont, Guy A. ; Sandor, George G.S. ; Potts, M.Terri ; Potts, James E.
Author_Institution :
Department of Electrical and Computer Engineering, the University of British Columbia, Vancouver, CANADA
fYear :
2008
fDate :
20-25 Aug. 2008
Firstpage :
5286
Lastpage :
5289
Abstract :
A novel method to obtain the aortic pressure waveform using a sequence of B-mode images is developed in this project. An automatic edge detection algorithm is applied to a sequence of longitudinal images of the aortic arch acquired from a suprasternal view. The aortic distension waveform is obtained by measuring the distance between the two edges throughout the cardiac cycle. It is then calibrated using the systolic and diastolic pressures from the brachial artery to obtain an estimated pressure waveform. This method was applied to 5 healthy children, pulse pressure amplification and total arterial compliance were calculated from the estimated waveforms.
Keywords :
Blood pressure; Brachytherapy; Cardiology; Carotid arteries; Image edge detection; Pediatrics; Pressure measurement; Pulse measurements; Ultrasonic imaging; Ultrasonic variables measurement; Adolescent; Algorithms; Aorta; Blood Pressure Determination; Child; Female; Humans; Image Enhancement; Image Interpretation, Computer-Assisted; Male; Pattern Recognition, Automated; Reproducibility of Results; Sensitivity and Specificity; Sternum; Ultrasonography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location :
Vancouver, BC
ISSN :
1557-170X
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2008.4650407
Filename :
4650407
Link To Document :
بازگشت