• DocumentCode
    3085536
  • Title

    Holographic Microwave Imaging Array: Experimental investigation of breast tumour detection

  • Author

    Lulu Wang ; Simpkin, Ray ; Al-Jumaily, A.M.

  • Author_Institution
    Inst. of Biomed. Technol., Auckland Univ. of Technol., Auckland, New Zealand
  • fYear
    2013
  • fDate
    1-3 Aug. 2013
  • Firstpage
    61
  • Lastpage
    64
  • Abstract
    This paper presents a two dimensional (2D) Holographic Microwave Imaging Array (HMIA) technique for early breast tumour detection and demonstrates its efficacy using experimental data. The system is designed for operation at a single frequency of 12.6 GHz, which includes one transmitter and an array of 15 receivers that is placed at a far-field under the breast phantom. The breast phantom consists of homogeneous normal breast tissue, a small malignant tumour and skin. The matching solution medium is not necessary in this system. Air is used between the antennas and breast phantom. The experimental results showed that small tumours (<;5 mm) at different locations could be successfully detected by using the proposed 2D HMIA technique.
  • Keywords
    biomedical imaging; holography; microwave imaging; receivers; transmitters; tumours; 2D HMIA technique; breast phantom; breast tumour detection; frequency 12.6 GHz; holographic microwave imaging array; malignant tumour; matching solution medium; receivers; skin; transmitter; two dimensional holographic microwave imaging array technique; Antennas; Arrays; Artificial intelligence; Biomedical measurement; Breast; Imaging; Tumors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetics (iWEM), 2013 IEEE International Workshop on
  • Conference_Location
    Kowloon
  • Print_ISBN
    978-1-4799-6654-7
  • Type

    conf

  • DOI
    10.1109/iWEM.2013.6888770
  • Filename
    6888770