Abstract :
We extended the concept of Short-lag Spatial Coherence imaging to matrix arrays and used it to improve the quality of vessel images. Specifically, we collected signals from the individual channels of a clinical matrix array while imaging a flow phantom and the human livers in vivo. We used this data to reconstruct matched B-mode and SLSC volumes. 2-D and 3-D images created from the SLSC volumes showed suppressed clutter and better visibility of vessels compared to their B-mode counterparts. Vessel contrast and CNR were computed for the volumes using concurrently acquired Doppler data as a gold standard for vessel visualization.
Keywords :
Doppler measurement; biomedical ultrasonics; blood; blood flow measurement; blood vessels; data acquisition; flow visualisation; image matching; image reconstruction; liver; medical image processing; phantoms; ultrasonic imaging; 2D images; 3D images; Doppler data acquisition; clinical matrix array; flow phantom; human livers in vivo; matched B-mode reconstruction; short-lag spatial coherence imaging; signal collection; vasculature; vessel image quality; vessel visualization; volumetric SLSC imaging; Arrays; Biomedical imaging; Clutter; Doppler effect; In vivo; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2013 IEEE International
Conference_Location :
Prague
ISSN :
1948-5719
Print_ISBN :
978-1-4673-5684-8
Type :
conf
DOI :
10.1109/ULTSYM.2013.0317
Filename :
6725307
Link To Document :
بازگشت