Title :
Comparison of control strategies for an EMG controlled orthotic exoskeleton for the hand
Author :
DiCicco, Matthew ; Lucas, Lenny ; Matsuoka, Yoky
Author_Institution :
Dept. of Mech. Eng., Carnegie Mellon Univ., Pittsburgh, PA, USA
fDate :
April 26-May 1, 2004
Abstract :
To restore dexterity to paralyzed hands, we have designed and constructed a lightweight, low-profile orthotic exoskeleton controlled by the user´s residual electromyography (EMG) signals. In this paper, we compared several simple strategies to control the orthotic device for a quadriplegic (C5/C6) subject. When contralateral arm control was employed, we found that a simple on/off strategy allowed for faster interaction with objects, while variable control provided more controlled interactions, especially with deformable objects. Furthermore, we designed a control strategy that allowed for a natural reaching and pinching sequence without the use of the contralateral arm. We validated that the EMG signal from the ipsilateral biceps could be used to develop an extremely reliable natural reaching and pinching algorithm. This evaluation showed that different control strategies may be appropriate for different situations, and further investigation on the natural algorithm is crucial.
Keywords :
control system synthesis; electromyography; medical control systems; medical signal processing; orthopaedics; orthotics; EMG controlled orthotic exoskeleton; contralateral arm control; electromyography signals; ipsilateral biceps; natural reaching sequence; orthotic device control; paralyzed hands; pinching sequence; quadriplegic subject; Design engineering; Electric variables control; Electromyography; Exoskeletons; Fasteners; Feedback; Fingers; Muscles; Orthotics; Signal restoration;
Conference_Titel :
Robotics and Automation, 2004. Proceedings. ICRA '04. 2004 IEEE International Conference on
Print_ISBN :
0-7803-8232-3
DOI :
10.1109/ROBOT.2004.1308056