Abstract
A dual-rate Kalman filter is developed for realtime continuous glucose monitoring. Frequent (5 minute) sampling of a noisy, continuous glucose sensor is used for estimation of glucose and its rate-of-change. Infrequent (8 hour intervals) reference glucose meter samples enable the sensor gain and its rate-of-change to be updated. The dual-rate Kalman filter formulation accounts for uncertainty in both the continuous glucose sensor and the reference glucose meter. The method is tested on simulated and experimental data, confirming its superiority to simple one-point calibration
Keywords
Kalman filters; biochemistry; biomedical measurement; blood; chemical sensors; 5 min; blood glucose sensor; dual-rate Kalman filter; glucose meter; realtime continuous glucose monitoring; sampling; sensor gain; sensor rate-of-change; Blood; Calibration; Cardiac disease; Chemical engineering; Degradation; Diabetes; Monitoring; Sampling methods; Sensor systems; Sugar;