• DocumentCode
    3109700
  • Title

    Selective characteristic for needle electrode arrays for functional electrical stimulation

  • Author

    Zhao, Xin-Tai ; Wang, Zhi-Gong ; Lü, Xiao-Ying ; Jiang, Zheng-Lin ; Lü, Guang-Ming ; Shen, Xiao-Yan ; Li, Wen-Yuan ; Huang, Zhong-Hao

  • Author_Institution
    Inst. of RF& OE-ICs, Southeast Univ., Nanjing, China
  • fYear
    2011
  • fDate
    26-28 March 2011
  • Firstpage
    1382
  • Lastpage
    1384
  • Abstract
    Selective characteristic for functional electrical stimulation is very important for restoring lost motor or sensory functions. The stimulation should broadly generate action potentials with a stimulating electrode and simultaneously blocks unnecessary stimulation with given grounded ones. 1-, 2-, 3-, and 5-needle electrode arrays were designed and simulated with finite element software, respectively. Animal experiments were carried out and the validity of the numerical results is demonstrated. Activating function were computed in the aforementioned four cases. The results show that the array with only one grounded electrode provides the best selective activation in the four cases.
  • Keywords
    bioelectric phenomena; biomedical electrodes; neurophysiology; finite element software; functional electrical stimulation; lost motor functions; lost sensory functions; needle electrode arrays; selective characteristics; Animals; Electric fields; Electric potential; Electrodes; Needles; Pain; Spinal cord;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Information Science and Technology (ICIST), 2011 International Conference on
  • Conference_Location
    Nanjing
  • Print_ISBN
    978-1-4244-9440-8
  • Type

    conf

  • DOI
    10.1109/ICIST.2011.5765094
  • Filename
    5765094