DocumentCode
3109700
Title
Selective characteristic for needle electrode arrays for functional electrical stimulation
Author
Zhao, Xin-Tai ; Wang, Zhi-Gong ; Lü, Xiao-Ying ; Jiang, Zheng-Lin ; Lü, Guang-Ming ; Shen, Xiao-Yan ; Li, Wen-Yuan ; Huang, Zhong-Hao
Author_Institution
Inst. of RF& OE-ICs, Southeast Univ., Nanjing, China
fYear
2011
fDate
26-28 March 2011
Firstpage
1382
Lastpage
1384
Abstract
Selective characteristic for functional electrical stimulation is very important for restoring lost motor or sensory functions. The stimulation should broadly generate action potentials with a stimulating electrode and simultaneously blocks unnecessary stimulation with given grounded ones. 1-, 2-, 3-, and 5-needle electrode arrays were designed and simulated with finite element software, respectively. Animal experiments were carried out and the validity of the numerical results is demonstrated. Activating function were computed in the aforementioned four cases. The results show that the array with only one grounded electrode provides the best selective activation in the four cases.
Keywords
bioelectric phenomena; biomedical electrodes; neurophysiology; finite element software; functional electrical stimulation; lost motor functions; lost sensory functions; needle electrode arrays; selective characteristics; Animals; Electric fields; Electric potential; Electrodes; Needles; Pain; Spinal cord;
fLanguage
English
Publisher
ieee
Conference_Titel
Information Science and Technology (ICIST), 2011 International Conference on
Conference_Location
Nanjing
Print_ISBN
978-1-4244-9440-8
Type
conf
DOI
10.1109/ICIST.2011.5765094
Filename
5765094