DocumentCode
3111859
Title
Noninvasive measurement of total hemoglobin and hemoglobin derivatives using multiwavelength pulse spectrophotometry -In vitro study with a mock circulatory system
Author
Suzaki, Hironori ; Kobayashi, Naoki ; Nagaoka, Takashi ; Iwasaki, Kiyotaka ; Umezu, Mitsuo ; Takeda, Sunao ; Togawa, Tatsuo
Author_Institution
Graduated Sch. of Sci. & Eng., Waseda Univ., Tokyo
fYear
2006
fDate
Aug. 30 2006-Sept. 3 2006
Firstpage
799
Lastpage
802
Abstract
Total hemoglobin (tHb), carboxyhemoglobin (COHb), and methemoglobin (MetHb) are usually measured with a CO-oximeter. Noninvasive and continuous measurement of these blood components is expected to decrease the pain of a patient. Therefore, we developed an instrument to measure oxygen saturation (SpO2), tHb, COHb, and MetHb non invasively. Multiwavelength LED (600, 625, 660, 760, 800, 940, and 1300 nm) and a combined detector (Si, InGaAs) were built into the instrument (Seven wavelengths transparent pulse spectrophotometer). We used the Waseda mock circulatory system, which can simulate blood circulation in tissues and generate a pulse wave mechanically, to estimate the instrument´s performance. Furthermore we proposed new calculation formula including DC components of optical density (this method). Under conditions without any change of other components, the mean error plusmn standard deviation between this method and the CO-oximetry were SaO2=0.0plusmn1.4%, tHb=0.0plusmn0.0 g/dl, COHb=0.0plusmn2.0%, and MetHb=0.0plusmn0.3%. When the concentration of other components was changed, this method showed mean errors and standard deviations of SaO2=0.2plusmn1.6%, tHb=0.0plusmn0.4 g/dl, COHb=0.5plusmn4.1%, and MetHb=0.0plusmn0.3%
Keywords
biochemistry; biological tissues; blood; haemodynamics; light emitting diodes; molecular biophysics; oximetry; proteins; spectrophotometry; 1300 nm; 600 nm; 625 nm; 660 nm; 760 nm; 800 nm; 940 nm; CO-oximetry; InGaAs; Si; Waseda mock circulatory system; blood circulation simulation; carboxyhemoglobin; methemoglobin; multiwavelength LED; multiwavelength pulse spectrophotometry; noninvasive measurement; optical density; oxygen saturation measurement; tissue; total hemoglobin derivatives; Blood; Circulatory system; Detectors; Instruments; Light emitting diodes; Noninvasive treatment; Optical pulse generation; Pain; Pulse measurements; Wavelength measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
Conference_Location
New York, NY
ISSN
1557-170X
Print_ISBN
1-4244-0032-5
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2006.260273
Filename
4461872
Link To Document