Title :
Biosensor multiwell microplate with self-reference capability
Author :
Cunningham, B. ; Baird, C. ; Li, P. ; Schulz, S. ; Albertson, K. ; Gerstenmaier, J.
Author_Institution :
SRU Biosystems, Woburn, MA, USA
Abstract :
The ability to remove sources of measurement error accurately and easily is one of the keys to a high-resolution biodetection system. The ability of a guided mode resonant filter (GMRF) biosensor structure to confine the lateral propagation of photons tightly is exploited to demonstrate a sensor surface upon which active and reference regions can be located in close physical proximity. Detection systems with two-point reference correction and an imaging-based reference method are demonstrated.
Keywords :
biosensors; measurement errors; microsensors; optical filters; optical information processing; optical sensors; biosensor multiwell microplate; guided mode resonant filter biosensor structure; high-resolution biodetection system; imaging-based reference method; lateral photon propagation; measurement error source removal; optical biosensors; self-reference capability; two-point reference correction; Biosensors; Error correction; Image sensors; Optical sensors; Optical surface waves; Optoelectronic and photonic sensors; Pixel; Resonance; Surface waves; Temperature sensors;
Conference_Titel :
Sensors, 2004. Proceedings of IEEE
Print_ISBN :
0-7803-8692-2
DOI :
10.1109/ICSENS.2004.1426291