• DocumentCode
    3122666
  • Title

    The development of non-sticking, passively cooled electrosurgical instruments

  • Author

    Sha, David ; Connolly, Raymond ; Schwaitzberg, Steven ; Levine, Andy

  • Author_Institution
    Seedling Enterprises, Newton, MA, USA
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    11
  • Lastpage
    12
  • Abstract
    An undesired side effect of electrosurgery includes tissue charring, smoke generation and the adhesion of tissue to electrodes. These effects prolong surgery and interfere with effective coagulation. New instruments are described that remove heat from the tips thereby maintaining their temperature below the sticking temperature. These cooled instruments cauterize normally but eliminate sticking and reduce char and smoke formation. These instruments are fabricated using heat pipes that passively dissipate heat along their length via natural convection. Infrared imaging is used to characterize performance and optimize electrode heat transfer surface area. In-vivo animal testing demonstrates excellent haemostasis while eliminating sticking. These new instruments may speed surgery and enable new applications for electrocautery
  • Keywords
    bioelectric phenomena; biomedical equipment; cooling; surgery; electrocautery; haemostasis; heat pipes; in-vivo animal testing; infrared imaging; medical instruments; natural convection; nonsticking passively cooled electrosurgical instruments; smoke generation; tissue adhesion to electrodes; tissue charring; Adhesives; Animals; Coagulation; Electrodes; Heat transfer; Infrared heating; Infrared imaging; Instruments; Surgery; Temperature;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioengineering Conference, 2001. Proceedings of the IEEE 27th Annual Northeast
  • Conference_Location
    Storrs, CT
  • Print_ISBN
    0-7803-6717-0
  • Type

    conf

  • DOI
    10.1109/NEBC.2001.924695
  • Filename
    924695