Title :
Cuffless Estimation of Systolic Blood Pressure for Short Effort Bicycle Tests: The Prominent Role of the Pre-Ejection Period
Author :
Muehlsteff, J. ; Aubert, X.L. ; Schuett, M.
Author_Institution :
Philips Res. Lab. Europe, Aachen
fDate :
Aug. 30 2006-Sept. 3 2006
Abstract :
This paper investigates the specific contributions of the pre-ejection period (PEP) and pulse transit time (PTT) for blood pressure estimation based on the pulse wave methodology. We show that in short-term physical stress tests, PEP dominates PTT variations raising the question of a suitable blood pressure calibration. A model using a generalized pulse wave velocity achieves acceptable accuracy for systolic blood pressure estimation, given our experimental conditions
Keywords :
biomechanics; blood pressure measurement; calibration; patient monitoring; bicycle test; blood pressure calibration; cuffless estimation; generalized pulse wave velocity; preejection period; pulse transit time; pulse wave methodology; short-term physical stress exercise; systolic blood pressure estimation; Bicycles; Biomedical monitoring; Blood pressure; Calibration; Electrocardiography; Fingers; Hypertension; Pressure measurement; Pulse measurements; Testing; blood pressure calibration; blood pressure regulation; physical stress; pre-ejection period; pulse transit-time;
Conference_Titel :
Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
Conference_Location :
New York, NY
Print_ISBN :
1-4244-0032-5
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2006.260275