DocumentCode :
3134170
Title :
The response of electric field probes to realistic RF environments
Author :
Adamson, David ; Bownds, Daniel ; Fernández, Alva ; Goodall, Edward
Author_Institution :
Nat. Phys. Lab., Teddington, UK
fYear :
2010
fDate :
23-28 May 2010
Firstpage :
1592
Lastpage :
1595
Abstract :
E-field probes used for measurement of worker exposure are usually calibrated in continuous wave (CW) conditions. For many types of probe, the response in realistic environments (modulated signals, multi-frequency and pulsed signals) differs from the response to CW conditions. However, to fully characterize a field probe in all possible environments would be an infinite task. In this paper we consider the effect of multiple CW signals and of several common communications formats on the indication given by a field probe with the objective of identifying simple tests which could act as a predictor of the performance alteration in non-CW environments.
Keywords :
electric field measurement; probes; CW signal; E-field probe; continuous wave signal; electric field measurement; electric field probe; Calibration; Diodes; Electric variables measurement; Envelope detectors; Frequency measurement; Frequency modulation; Probes; Pulse modulation; Radio frequency; Testing; Electric field measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Symposium Digest (MTT), 2010 IEEE MTT-S International
Conference_Location :
Anaheim, CA
ISSN :
0149-645X
Print_ISBN :
978-1-4244-6056-4
Electronic_ISBN :
0149-645X
Type :
conf
DOI :
10.1109/MWSYM.2010.5517119
Filename :
5517119
Link To Document :
بازگشت