• DocumentCode
    3135878
  • Title

    Measurement of Residual Solvents in Thiamine Hydrochloride by Static Headspace Gas Chromatography

  • Author

    Song, Wenhua ; Gao, Minling ; Guo, Jing ; Sun, Ting

  • Author_Institution
    Sch. of Environ. & Chem. Eng., Tianjin Polytech. Univ., Tianjin, China
  • fYear
    2010
  • fDate
    18-20 June 2010
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A rapid general method for the determination of residual solvents in thiamine was studied. This paper demonstrated static headspace gas chromatography (HS/GC) with either flame ionization detection (FID) or mass spectrometry (MS) for the determination of volatile residual solvents in pharmaceutical products. The analytical method performance characteristics were investigated, including the repeatability, the accuracy and the detection limit of determination. The results showed that the method had very good sensitivities (detection limits under 9.45 μg/mL) and precision (under 7.35% RSD) for all solvents. The rates of recovery were 98.59%, 95.25%, 103.89% and 99.88% for methanol, ethanol, N, N-dimethyllformamide and ethyl ether, respectively.
  • Keywords
    chromatography; mass spectroscopy; pharmaceuticals; N, N-dimethyllformamide; ethanol; ethyl ether; flame ionization detection; mass spectrometry; methanol; pharmaceutical products; residual solvents; static headspace gas chromatography; thiamine hydrochloride; volatile residual solvents; Drugs; Ethanol; Fires; Gas chromatography; Linearity; Methanol; Pharmaceuticals; Sampling methods; Solvents; Temperature;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioinformatics and Biomedical Engineering (iCBBE), 2010 4th International Conference on
  • Conference_Location
    Chengdu
  • ISSN
    2151-7614
  • Print_ISBN
    978-1-4244-4712-1
  • Electronic_ISBN
    2151-7614
  • Type

    conf

  • DOI
    10.1109/ICBBE.2010.5517221
  • Filename
    5517221