DocumentCode :
3137163
Title :
Using AC signal and transimpedance amplifier circuit to build a glucose biosensor model
Author :
Sun, Tai-Ping ; Shieh, Hsiu-Li ; Yeh, Wei-Cheng ; Tang, Yi-Ching ; Tsai, Chun-Lang ; Huang, Hsin-Wei ; Liu, Chia-Ming ; Chang, Jen-Wei ; Li, Jia-Hao ; Ching, Congo Tak-Shing
Author_Institution :
Grad. Inst. of Biomedicine & Biomed. Technol., Nat. Chi Nan Univ., Nantou Hsien, Taiwan
Volume :
5
fYear :
2010
fDate :
16-18 Oct. 2010
Firstpage :
2142
Lastpage :
2144
Abstract :
In this study, a glucose biosensor model was developed. The experiment was performed by connecting commercial amperiometric glucose biosensor with transimpedance amplifier (TIA) readout circuit. By supplying various AC signals and glucose solutions to the biosensor, the resistance and capacitance of the biosensor was found. In conclusion, an amperiometric glucose biosensor model is a parallel combination of 2 resistors (1 for biosensor and 1 for enzymatic reaction) and 2 capacitors (1 for biosensor and 1 for enzymatic reaction).
Keywords :
amperometric sensors; biosensors; catalysis; enzymes; molecular biophysics; operational amplifiers; AC signal; amperometric glucose biosensor; biosensor capacitance; biosensor resistance; enzymatic reaction; glucose biosensor model; transimpedance amplifier readout circuit; Biological system modeling; Biosensors; Capacitance; Immune system; Integrated circuit modeling; Mathematical model; Sugar; Amperometric Glucose biosensor; Transimpedance amplifier;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Engineering and Informatics (BMEI), 2010 3rd International Conference on
Conference_Location :
Yantai
Print_ISBN :
978-1-4244-6495-1
Type :
conf
DOI :
10.1109/BMEI.2010.5639281
Filename :
5639281
Link To Document :
بازگشت