DocumentCode
3146058
Title
Flexible Parylene Packaged Intraocular Coil for Retinal Prostheses
Author
Li, Wen ; Rodger, Damien C. ; Meng, Ellis ; Weiland, James D. ; Humayun, Mark S. ; Tai, Yu-Chong
Author_Institution
California Inst. of Technol., Pasadena, CA
fYear
2006
fDate
9-12 May 2006
Firstpage
105
Lastpage
108
Abstract
We present a flexible and fully-implantable coil designed for use as a power and data transfer component in retinal prosthesis applications. Compared with traditional hand-made intraocular coils, this microfabricated coil is flexible, with a 9.5 mm outer diameter and 10-mum-thick parylene C as the primary structural and packaging material. A post-fabrication heat treatment was used to improve the parylene package in order to protect the device in harsh corrosive environments such as the human eye. Long-term accelerated-lifetime soak testing in heated saline has been performed, and the mean time to failure (MTTF) of the parylene package extrapolated to 37degC was estimated using the Arrhenius relationship. The electrical failure of this device was also characterized by measuring the DC resistance in saline
Keywords
bioMEMS; coils; eye; prosthetics; 10 micron; 37 degC; 9.5 mm; Arrhenius relationship; DC resistance; data transfer component; flexible parylene packaged intraocular coil; heat treatment; microfabricated coil; packaging material; parylene C; retinal prostheses; Coils; Heat treatment; Humans; Life estimation; Packaging; Performance evaluation; Prosthetics; Protection; Retina; Testing; Accelerated-lifetime soak testing; Coil; Intraocular; Parylene; Retinal prostheses;
fLanguage
English
Publisher
ieee
Conference_Titel
Microtechnologies in Medicine and Biology, 2006 International Conference on
Conference_Location
Okinawa
Print_ISBN
1-4244-0338-3
Electronic_ISBN
1-4244-0338-3
Type
conf
DOI
10.1109/MMB.2006.251502
Filename
4281320
Link To Document