• DocumentCode
    3146058
  • Title

    Flexible Parylene Packaged Intraocular Coil for Retinal Prostheses

  • Author

    Li, Wen ; Rodger, Damien C. ; Meng, Ellis ; Weiland, James D. ; Humayun, Mark S. ; Tai, Yu-Chong

  • Author_Institution
    California Inst. of Technol., Pasadena, CA
  • fYear
    2006
  • fDate
    9-12 May 2006
  • Firstpage
    105
  • Lastpage
    108
  • Abstract
    We present a flexible and fully-implantable coil designed for use as a power and data transfer component in retinal prosthesis applications. Compared with traditional hand-made intraocular coils, this microfabricated coil is flexible, with a 9.5 mm outer diameter and 10-mum-thick parylene C as the primary structural and packaging material. A post-fabrication heat treatment was used to improve the parylene package in order to protect the device in harsh corrosive environments such as the human eye. Long-term accelerated-lifetime soak testing in heated saline has been performed, and the mean time to failure (MTTF) of the parylene package extrapolated to 37degC was estimated using the Arrhenius relationship. The electrical failure of this device was also characterized by measuring the DC resistance in saline
  • Keywords
    bioMEMS; coils; eye; prosthetics; 10 micron; 37 degC; 9.5 mm; Arrhenius relationship; DC resistance; data transfer component; flexible parylene packaged intraocular coil; heat treatment; microfabricated coil; packaging material; parylene C; retinal prostheses; Coils; Heat treatment; Humans; Life estimation; Packaging; Performance evaluation; Prosthetics; Protection; Retina; Testing; Accelerated-lifetime soak testing; Coil; Intraocular; Parylene; Retinal prostheses;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microtechnologies in Medicine and Biology, 2006 International Conference on
  • Conference_Location
    Okinawa
  • Print_ISBN
    1-4244-0338-3
  • Electronic_ISBN
    1-4244-0338-3
  • Type

    conf

  • DOI
    10.1109/MMB.2006.251502
  • Filename
    4281320