DocumentCode :
3161827
Title :
Development of a gas flow calibration facility by master meter method
Author :
Shengjie, Li ; Lijun, Sun ; Lixiao, Qi ; Lei, Qi ; Tao, Zhang
Author_Institution :
Tianjin Key Lab. of Process Meas. & Control, Tianjin Univ., Tianjin, China
fYear :
2011
fDate :
16-18 April 2011
Firstpage :
2126
Lastpage :
2129
Abstract :
In this paper, a gas flow calibration facility by master meter method is developed. The paper deals with the design methods and working principle. The simulation of the confluence tube is analyzed. The pressure loss of pipe is calculated and the selection of the blower is analyzed. And then the distribution of the facility uncertainty is calculated. The uncertainty is better than 1% in the total flow range, and it is better than 0.5% at the fixed flow points.
Keywords :
calibration; design engineering; flow measurement; gas industry; pipe flow; pipes; blower selection; confluence tube; design method; gas flow calibration facility; master meter method; pipe; pressure loss; Calibration; Electron tubes; Fluid flow; Fluid flow measurement; Instruments; Publishing; Uncertainty; blower sections; confluence tube; flow measurement; gas flow calibration facility; pipe pressure loss; uncertainty;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Consumer Electronics, Communications and Networks (CECNet), 2011 International Conference on
Conference_Location :
XianNing
Print_ISBN :
978-1-61284-458-9
Type :
conf
DOI :
10.1109/CECNET.2011.5768934
Filename :
5768934
Link To Document :
بازگشت