Title :
Effect of contrast injection on coronary artery pressure and dimension during angiography
Author :
Stiel, Georg M. ; Lattermann, Andreas ; Schaps, Klaus-Peter ; Siebes, Maria ; Nienaber, Christoph A.
Author_Institution :
Dept. of Internal Med., Univ. Hospital Eppendorf, Hamburg, Germany
Abstract :
Digital coronary angiography and intracoronary pressure recordings were performed in 17 patients to quantitate the mechanical effects of contrast injection on coronary artery pressure and diameter in arterial segments free of stenotic lesions at the moment of image acquisition. A progressive increase in intracoronary pressures by 9.2±6.6 mmHg for a power injection rate of 2.5 ml/s, 14.4±10.9 mmHg for 5 ml/s and 23.9±24.2 mmHg for 10 ml/s was documented Corresponding diameters were distended by net -0.13 to +0.65 mm (p<0.005) during the injection, as evaluated by digital quantitative coronary angiography (DQCA). These mechanical phenomena result in changes of coronary artery, diameter from -4.9% to +19.5% at 10 ml/s (p<0.005) and unmask the individual compliance of the arterial wall. It is concluded that coronary dimensions obtained from DQCA critically depend on the intracoronary pressure with contrast medium injection. It may represent an additional source of error in the analysis and interpretation of clinical angiographic studies
Keywords :
angiocardiography; blood pressure measurement; diagnostic radiography; diameter measurement; 23.9 mmHg; 9.2 mmHg; arterial wall compliance; contrast injection; contrast injection effect; coronary artery dimension; coronary artery pressure; digital coronary angiography; intracoronary pressure recording; mechanical effects quantification; stenotic lesions; Angiography; Arteries; Biomedical imaging; Cardiology; Clinical trials; Digital recording; Image segmentation; Lesions; Pressure measurement; Quantum cellular automata;
Conference_Titel :
Computers in Cardiology 1995
Conference_Location :
Vienna
Print_ISBN :
0-7803-3053-6
DOI :
10.1109/CIC.1995.482596