DocumentCode
3173990
Title
Ejection fraction with planar gated 99m-Tc SestaMIBI images
Author
Szulc, M. ; Herrold, EM ; Zanzonico, P. ; Borer, JS
Author_Institution
Cornell Univ. Med. Center, New York, NY, USA
fYear
1995
fDate
10-13 Sept. 1995
Firstpage
521
Lastpage
524
Abstract
Left ventricular (LV) ejection fraction (EF) can be determined from planar first pass blood pool images following injection of 99m-Tc SestaMIBI. However, it is technically simpler to obtain LVEF from gated SestaMIBI images of the perfused myocardium. Because the geometry-based EF calculation involves potential errors, we assessed the accuracy of this method by both validating the approach using phantom images, and comparing LVEFs from perfusion studies with the LVEFs from contrast angiography in the same patients. We also demonstrated that this bidimensional EF calculation method reflects changes in LV volumes from end diastole (ED) to end systole (ES), as in an ellipsoid model of the left ventricle, by supporting the assumption that changes along the ellipsoid long axis from ED to ES are significantly smaller than changes along the anteroposterior or septum-lateral axes of the ventricle, thus making LVEF from planar gated perfusion images a useful diagnostic and prognostic tool.
Keywords
angiocardiography; electrocardiography; radioisotope imaging; ECG gated myocardial perfusion planar images; LV ejection fraction; Tc; accuracy; anteroposterior axes; bidimensional EF calculation method; contrast angiography; diagnostic tool; ellipsoid long axis; ellipsoid model; end diastole; end systole; errors; first pass cineangiography; geometry-based EF calculation; left ventricle; left ventricular ejection fraction; perfused myocardium; phantom images; planar first pass blood pool images; planar gated 99m-Tc SestaMIBI images; planar gated perfusion images; prognostic tool; septum-lateral axes; Biomedical imaging; Blood; Cameras; Cardiology; Catheterization; Collimators; Ellipsoids; Imaging phantoms; Medical diagnostic imaging; Myocardium;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology 1995
Conference_Location
Vienna, Austria
Print_ISBN
0-7803-3053-6
Type
conf
DOI
10.1109/CIC.1995.482716
Filename
482716
Link To Document