Title :
Quantitative coronary arteriography: improving catheter calibration by an efficient correction for differential magnification
Author :
Fischer, F. ; Wunderlich, W. ; Röhrig, B. ; Kristiansen, G. ; Linderer, T. ; Schultheiss, HP
Author_Institution :
Dept. of Cardiology, Freie Univ. Berlin, Germany
Abstract :
Differential magnification of the vessel and of the scaling catheter in the X-ray field causes significant errors in calibration of imaged vessel size. To correct for differential magnification properly, laborious and costly calculations are required. The authors derived the equations of accurate correction for differential magnification. Starting from these equations the authors reduced the correction effort to the acquisition of a second complementary angiogram and two digitizing operations. The validation of this approximate procedure in computer simulations and in vitro phantom studies showed a correction for differential magnification up to 85%.
Keywords :
angiocardiography; calibration; diagnostic radiography; blood vessel; catheter calibration improvement; computer simulations; differential magnification; digitizing operations; efficient correction; imaged vessel size; in vitro phantom studies; medical diagnostic imaging; quantitative coronary arteriography; second complementary angiogram; Angiography; Calibration; Cardiology; Catheters; Computer simulation; Delta modulation; Equations; Error correction; Imaging phantoms; In vitro;
Conference_Titel :
Computers in Cardiology 1995
Conference_Location :
Vienna, Austria
Print_ISBN :
0-7803-3053-6
DOI :
10.1109/CIC.1995.482724