Title :
Comparison of fusion techniques for 3D+T echocardiography acquisitions from different acoustic windows
Author :
Soler, P. ; Gérard, O. ; Allain, Pascal ; Saloux, E. ; Angelini, E. ; Bloch, I.
Author_Institution :
Philips Med. Syst. Res., Paris
Abstract :
We propose a new method to combine real-time 3D echocardiography acquisitions from different apical windows. It consists in registering the volumes using only the image data, without external positioning sensors, and combining their intensity values, without requiring a feature extraction method. Registration shows reasonable quantitative results on phantom data and visually acceptable results on in vivo data. We present two new methods for the fusion process: generalized averaging and multiview deconvolution. The latter, which performed best, improved the myocardium contrast by +49.67% and the signal-to-noise ratio by +4.61 dB. These improvements lead to more accurate quantification of cardiac function
Keywords :
acoustic materials; cardiovascular system; deconvolution; echocardiography; image registration; medical image processing; sensor fusion; 3D+T echocardiographic acquisition; acoustic apical windows; cardiac function quantification; fusion techniques; generalized averaging deconvolution; image data; multiview deconvolution; myocardium contrast; phantom data; real-time 3D echocardiography; signal-to-noise ratio; volume registration; Echocardiography;
Conference_Titel :
Computers in Cardiology, 2005
Conference_Location :
Lyon
Print_ISBN :
0-7803-9337-6
DOI :
10.1109/CIC.2005.1588054