• DocumentCode
    3174344
  • Title

    Studies on cantilever based triglyceride biosensor

  • Author

    Fernandez, Renny Edwin ; Soma Sekhar, B.V. ; Bhattacharya, Enakshi ; Chadha, Anju

  • Author_Institution
    Indian Inst. of Technol. Madras, Chennai
  • fYear
    2007
  • fDate
    16-20 Dec. 2007
  • Firstpage
    690
  • Lastpage
    693
  • Abstract
    We report detection of micromolar levels of triglycerides using surface micromachined polysilicon cantilever beams. Enzymatic hydrolysis of triglycerides produces glycerol which alters the viscosity and density of the solution. This affects the dynamic properties of cantilever beams immersed in the solution. The change in the resonance frequency of the cantilever beams in the solution is measured using Doppler Vibrometry and the concentration of triglyceride is determined by comparing with a predetermined calibration plot.
  • Keywords
    beams (structures); biosensors; cantilevers; microsensors; viscosity; Doppler vibrometry; cantilever based triglyceride biosensor; cantilever beams; enzymatic hydrolysis; glycerol; micromolar levels detection; surface micromachined polysilicon cantilever beams; viscosity; Biosensors; Chemicals; Frequency measurement; Laser beams; Monitoring; Resonance; Resonant frequency; Structural beams; Vibrometers; Viscosity; biosensor; microcantilever; resonance frequency; triglyceride; vibrometer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Physics of Semiconductor Devices, 2007. IWPSD 2007. International Workshop on
  • Conference_Location
    Mumbai
  • Print_ISBN
    978-1-4244-1728-5
  • Electronic_ISBN
    978-1-4244-1728-5
  • Type

    conf

  • DOI
    10.1109/IWPSD.2007.4472613
  • Filename
    4472613