Title :
Automating the determination of wave speed using the PU-loop method
Author :
Swalen, Mjp ; Khir, Aw
Author_Institution :
Sch. of Eng., Brunel Univ., Uxbridge
Abstract :
The PU-loop (pressure-velocity loop) is a method for determining wave speed and relies on the linear relationship between the pressure and velocity in the absence of reflected waves. This linearity of the PU-loop during early systole, which is directly related to wave speed, has always been established by eye. This paper presents a new technique that establishes this linearity and thus determining wave speed online. Pressure and flow were measured in the ascending aorta of 11 anesthetised dogs. The slope of the PU-loop, indicating wave speed was determined by eye and by using the new technique. The difference between the slopes of the two methods is in the order of 3%. The new technique is convenient and allows for the online assessment of wave speed, which could be used as a bedside tool for the assessment of arterial compliance
Keywords :
biomedical ultrasonics; blood flow measurement; blood pressure measurement; blood vessels; cardiovascular system; PU-loop method; arterial compliance; ascending aorta; early systole; flow measurement; pressure measurement; pressure-velocity loop; ultrasonic flow meter; wave speed determination; Arteries; Catheters; Dogs; Fluid flow measurement; Linearity; Pressure measurement; Probes; USA Councils; Valves; Velocity measurement;
Conference_Titel :
Computers in Cardiology, 2005
Conference_Location :
Lyon
Print_ISBN :
0-7803-9337-6
DOI :
10.1109/CIC.2005.1588099