Abstract :
A drift calibration technique for DC-coupled EOG (electrooculogram) systems is proposed. It assumes a non-linear relationship between EOG and eye angle and estimates the absolute eye angle by the EOG differences during saccade. Drift is calibrated every saccade without user´s explicit action, so it is especially suitable for long-term gaze input interfaces. An experiment confirms that it can estimate horizontal absolute eye angle with an error of about 5° in addition to accurate eye movement.
Keywords :
angular measurement; calibration; electro-oculography; medical signal processing; user interfaces; DC coupled EOG systems; EOG based gaze input interface; absolute eye angle estimation; automatic drift calibration; drift calibration technique; electrooculogram; saccade movement; Calibration; Electrodes; Electrooculography; Equations; Estimation; Mathematical model; Tracking;