DocumentCode :
3185214
Title :
BaroLoop: Using a multichannel cuff electrode and selective stimulation to reduce blood pressure
Author :
Plachta, D.T.T. ; Gierthmuehlen, M. ; Cota, O. ; Boeser, Fabian ; Stieglitz, T.
Author_Institution :
Dept. of Microsyst. Eng., Univ. of Freiburg, Freiburg, Germany
fYear :
2013
fDate :
3-7 July 2013
Firstpage :
755
Lastpage :
758
Abstract :
The therapy of refractory hypertension is an increasing problem for health care systems and a frontend in research in both pharmacology and neuroelectronic engineering. Overriding the baroreceptive information of afferent nerve fibers, originating from pressure sensors in the aortic arch, can trigger the baroreflex, a systemic control system that lowers the blood pressure (BP) almost instantaneously. Using a multichannel cuff electrode, wrapped around a rat vagal nerve, we were able to regulate the BP using selective, tripolar stimulation. The tripolar stimulation was sufficiently selective to not trigger any unwanted side effects like bradycardia or bradypnea. The BP was reduced best with charge balanced stimulation amplitudes of 1 mA and pulse duration of 0.3 ms. The stimulation frequency had only a mild influence on the effectiveness of the stimulation and did work best at 40 Hz. We found that the BP took up to five times the stimulation period to recover to the value prior to stimulation.
Keywords :
biomedical electrodes; diseases; health care; neurophysiology; patient treatment; BaroLoop; aortic arch; baroreflex; blood pressure reduction; bradycardia; bradypnea; charge balanced stimulation amplitudes; current 1 mA; frequency 40 Hz; health care systems; multichannel cuff electrode; nerve fibers; neuroelectronic engineering; pharmacology; pressure sensors; pulse duration; rat vagal nerve; refractory hypertension therapy; stimulation frequency; systemic control system; time 0.3 ms; tripolar stimulation; Baroreflex; Blood pressure; Educational institutions; Electrodes; Heart rate; Impedance; Rats;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location :
Osaka
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2013.6609610
Filename :
6609610
Link To Document :
بازگشت