DocumentCode :
3187106
Title :
Simplified high accuracy high bandwidth calibration of dielectric capacitors
Author :
Svetik, Z.
Author_Institution :
SIQ, Ljubljana, Slovenia
Volume :
3
fYear :
2001
fDate :
2001
Firstpage :
1445
Abstract :
A novel method for calibration of dielectric capacitors is proposed using a commercial high resolution and high bandwidth impedance meter as a comparison device and high bandwidth four terminal pair standard resistors. The advantage of the method is that no sophisticated nor custom-built equipment needs to be used to achieve high accuracy and high frequency bandwidth coverage. The method is simple to apply once the measurements and calculations are automated and the influencing parameters properly evaluated. The calculated measurement uncertainties are better than 1 ·10-4 both for capacitance and dissipation factor confirmed by the agreement found with results obtained through other procedures
Keywords :
calibration; capacitance measurement; capacitors; frequency response; measurement uncertainty; 20 Hz to 1 MHz; 4TP impedance meter; capacitance; dielectric capacitors; dissipation factor; frequency response; high bandwidth calibration; high bandwidth impedance meter; high resolution; measurement uncertainty; terminal pair standard resistors; Bandwidth; Calibration; Capacitance; Capacitors; Dielectric measurements; Error correction; Frequency; Impedance measurement; Instruments; Resistors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2001. IMTC 2001. Proceedings of the 18th IEEE
Conference_Location :
Budapest
ISSN :
1091-5281
Print_ISBN :
0-7803-6646-8
Type :
conf
DOI :
10.1109/IMTC.2001.929443
Filename :
929443