• DocumentCode
    3189922
  • Title

    Free placement trans-skull Doppler system with 1.0MHz array ultrasonic probe

  • Author

    Nakamura, Masato ; Kitamura, Yuri T. ; Yanagida, Toshio ; Kobashi, Syoji ; Kuramoto, Kei ; Hata, Yutaka

  • Author_Institution
    Grad. Sch. of Eng., Univ. of Hyogo, Himeji, Japan
  • fYear
    2010
  • fDate
    10-13 Oct. 2010
  • Firstpage
    1370
  • Lastpage
    1374
  • Abstract
    This paper describes a trans-skull ultrasonic system that measures the blood flow velocity through the brain´s blood vessel under skull. In this system, we use an ultrasonic array probe with the center frequency of 1.0MHz. The system determines the blood flow by Doppler effect. This Doppler effect is examined by the center of gravity shift in the frequency domain. We test the system in the condition of the water flow in silicon tube under the cow scapula. The experimental result shows that the system detects the flow velocity by Doppler effect and confirms the usefulness of the method under skull.
  • Keywords
    Doppler measurement; biomedical ultrasonics; blood flow measurement; blood vessels; brain; ultrasonic arrays; array ultrasonic probe; blood flow velocity; brain blood vessel; cow scapula; free placement trans-skull Doppler system; frequency 1.0 MHz; frequency domain; gravity shift; silicon tube; water flow; Acoustics; Noise; Noise measurement; Blood Vessel; Brain Imaging; Doppler Effect; Medical System; Transcranial Ultrasonic System;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Systems Man and Cybernetics (SMC), 2010 IEEE International Conference on
  • Conference_Location
    Istanbul
  • ISSN
    1062-922X
  • Print_ISBN
    978-1-4244-6586-6
  • Type

    conf

  • DOI
    10.1109/ICSMC.2010.5642453
  • Filename
    5642453