Method for location of puncture point guided by digital mammography image
Author :
Jian Wu ; Peng Gao ; Xiu-wang Li ; Yi-Bin Huang
Author_Institution :
Center of Biomed. Eng., Tsinghua Univ., Shenzhen, China
fYear :
2013
fDate :
3-7 July 2013
Firstpage :
2368
Lastpage :
2371
Abstract :
The main purpose of this study was to develop a method that can optimize the algorithm for puncture point calculation, and therefore improve the accuracy of X-ray guided breast biopsy. The proposed method is: first, select two guiding points; then, use the guiding points to construct X-ray cone-beams, so the joining section of the cone-beams can be used to determine the puncture target point. The method was verified by a phantom emulation, in which the calculated target-points were all found within the central part of the target lesion, and far away from the boarder of the lesion (change the X-ray tube angle by 15° would only cause a slight deviation no more than 1.4 mm), the accuracy is enough to fulfill the needs of biopsy operation. This study also found out that, the shorter the distance between the guiding point and the center of the lesion´s project, the nearer the calculated biopsy-point will be to the actual lesion center.
Keywords :
X-ray tubes; diagnostic radiography; mammography; phantoms; X-ray cone-beam construction; X-ray guided breast biopsy; X-ray tube angle; algorithm optimization; digital mammography image; distance 1.4 mm; phantom emulation; puncture point location; target lesion; Biopsy; Breast biopsy; Ground penetrating radar; Lesions; Mathematical model; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location :
Osaka
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2013.6610014
Filename :
6610014
Link To Document :
بازگشت