DocumentCode :
3193691
Title :
Establishing traceability of an electronic calibration unit using the NIST Microwave Uncertainty Framework
Author :
Jargon, Jeffrey A. ; Williams, Dylan F. ; Wallis, T. Mitch ; LeGolvan, Denis X. ; Hale, Paul D.
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
fYear :
2012
fDate :
22-22 June 2012
Firstpage :
1
Lastpage :
5
Abstract :
We present a method for establishing traceability of a commercial electronic calibration unit for vector network analyzers by characterizing the scattering parameters of its internal states with repeated multiline thru-reflect-line (TRL) calibrations and utilizing the NIST Microwave Uncertainty Framework to propagate uncertainties. With the electronic calibration unit characterized, we use it to calibrate a network analyzer, and characterize a number of verification devices with corresponding uncertainties. We also characterize the same verification devices using one of the previous multiline TRL calibrations, and compare results.
Keywords :
S-parameters; calibration; measurement uncertainty; microwave measurement; network analysers; NIST microwave uncertainty framework; electronic calibration unit traceability; repeated multiline TRL calibration; repeated multiline thru-reflect-line calibration; scattering parameter; uncertainty propagation; vector network analyzer; Calibration; Measurement uncertainty; Microwave theory and techniques; NIST; Scattering parameters; Uncertainty; Electronic calibration unit; traceability; uncertainty; vector network analyzer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Measurement Conference (ARFTG), 2012 79th ARFTG
Conference_Location :
Montreal, QC
Print_ISBN :
978-1-4673-1229-5
Electronic_ISBN :
978-1-4673-1230-1
Type :
conf
DOI :
10.1109/ARFTG79.2012.6291181
Filename :
6291181