Abstract :
Ultra-low field magnetic resonance imaging (ULF-MRI) has attracted attention because of its low running costs and minimum patient exposure. An optically pumped atomic magnetometer (OPAM) is a magnetic sensor with high sensitivity in the low frequency range, which does not require a cryogenic cooling system. In an effort to develop a ULF-MRI, we attempted to measure the free induction decay MR signals with an OPAM. We successfully detected the MR signals by combining an OPAM and a flux transformer, demonstrating the feasibility of the proposed system.
Keywords :
biomedical MRI; biomedical equipment; magnetic variables measurement; magnetometers; OPAM; ULF-MRI; flux transformer; free induction decay MR signal measurements; free induction decay MR signals; magnetic resonance imaging; magnetic sensor; optically pumped atomic magnetometer; ultralow field MRI; Coils; Laser excitation; Magnetic noise; Magnetic resonance imaging; Magnetic shielding; Magnetometers; Resonant frequency;