Development of a portable bridge-to-decision blood pump
Author
Yamane, Toshiyuki ; Kitamura, Kokoro
Author_Institution
Mech. Eng. Dept., Kobe Univ., Kobe, Japan
fYear
2013
fDate
3-7 July 2013
Firstpage
2728
Lastpage
2731
Abstract
We are developing an axial-flow pump with a cylindrical-impeller without airfoils. In the mock experiments of HA02 model a pressure of 13.3 kPa was obtained at a rotational speed of 12500 rpm and flow of 5L/min. The obtained pressure with HA02 was almost double than an airfoil-type impeller. The 2D analysis of hydrodynamic bearings for the pump revealed that a section with 3 or more arcs is stable with respect to angular position, and a minimum bearing gap of 100μm can be attained at a design bearing gap of 150 μm and at a groove depth of 100μm.
Keywords
biomedical materials; haemodynamics; hydrodynamics; impellers; prosthetics; pumps; HA02 model; airfoil-type impeller; angular position; arc; axial flow pump development; cylindrical impeller; depth 100 mum; distance 100 mum; distance 150 mum; hydrodynamic bearing 2D analysis; mock experiment; portable bridge-to-decision blood pump; pressure 13.3 kPa; rotational speed; Atmospheric modeling; Automotive components; Blood; Hydrodynamics; Impellers; Pumps; Velocity control;