DocumentCode
3200054
Title
Optical Design of Micro Spectrometer
Author
Danyang Xu ; Chenghua Sui ; Jianping Tong ; Hao Yang
Author_Institution
Coll. of Sci., Zhejiang Univ. of Technol., Hangzhou, China
fYear
2012
fDate
8-10 Dec. 2012
Firstpage
339
Lastpage
341
Abstract
Optical system is the core of spectrometer decides directly spectrometer´s volume and weight. It is also the key to miniaturize spectrometer. In this article, in order to solve the miniaturization problem, the choice of optical construction, the computation of optical elements and light path and the elimination of aberration have been discussed. The result shows that the micro optical system´s spectral range is 200-910 nm, resolution is about 1nm and the system volume is 62.5 ×72.5×17.5 mm, and the whole system has compact structure and small volume and meets the requirement of minisize.
Keywords
ultraviolet spectrometers; microspectrometer; miniaturize spectrometer; optical design; optical elements; Gratings; Lenses; Mirrors; Optical design; Optical imaging; Optical sensors; Optical variables measurement; blazing grating; coma Introduction (Heading 1); cross-asymmetric Czerny-Turner system; spherical aberration;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation, Measurement, Computer, Communication and Control (IMCCC), 2012 Second International Conference on
Conference_Location
Harbin
Print_ISBN
978-1-4673-5034-1
Type
conf
DOI
10.1109/IMCCC.2012.83
Filename
6428916
Link To Document