An automatic capture control and threshold tracking system would be an important safety feature for the patient being thrown onto pacemaker therapy. In this connection the aim of this study is to investigate suitability of the VER as signal for such an application. Following this an algorithm was developed and checked on its reliability. These investigations performed by an external pacemaker system during 10 implantations show, that the VER is a preferable signal for automatic capture control and that the developed algorithm works reliably, The results of these investigations have been lead to the development of a new implantable pacemaker providing the capture control algorithm. First Investigations using this pacemaker in 8 patients show a safe detection of effective and not effective stimulation
Keywords
medical signal processing; pacemakers; safety; algorithm; automatic capture control; automatic capture recognition; capture control algorithm; effective stimulation; external pacemaker system; implantable pacemaker; pacemaker pulses; pacemaker therapy; patients; reliability; safe detection; threshold tracking system; ventricular evoked response; Automatic control; Biomedical engineering; Control systems; Electrodes; Energy consumption; Pacemakers; Pulse measurements; Safety; Surface discharges; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1996. Bridging Disciplines for Biomedicine. Proceedings of the 18th Annual International Conference of the IEEE
Conference_Location
Amsterdam
Print_ISBN
0-7803-3811-1
Type
conf
DOI
10.1109/IEMBS.1996.652812
Filename
652812
Link To Document