DocumentCode :
3205030
Title :
An Ultra Low Power Pulse Oximeter Sensor Based on Compressed Sensing
Author :
Baheti, Pawan K. ; Garudadri, Harinath
Author_Institution :
Qualcomm Inc., San Diego, CA, USA
fYear :
2009
fDate :
3-5 June 2009
Firstpage :
144
Lastpage :
148
Abstract :
We describe an ultra low power pulse oximeter sensor for long term, non-invasive monitoring of SpO2 and heart rate in Body Area Networks (BAN). Commercial pulse oximeter sensors consume about 20-60 mW of power during continuous operation. Other researchers have shown that accurate and noise robust wireless pulse oximeter sensors can be designed to operate with as little as 1.5 mW. The LEDs consume bulk of the power budget in pulse oximeter sensors. In this work, we describe a compressed sensing approach to sample the photodetector output, so that the LEDs can be turned off for longer periods and thus save sensor power. We randomly sample Photoplethysmogram (PPG) signals with about 10-40x fewer samples than with uniform sampling and demonstrate that the accuracy of heart rate estimation and blood pressure estimation are not compromised, using MIMIC database. This provides power savings of the order of 10-40x for a pulse oximeter sensor, by reducing the duration LEDs need to be turned on.
Keywords :
biomedical electronics; blood pressure measurement; body area networks; cardiology; light emitting diodes; low-power electronics; oximetry; photodetectors; plethysmography; LED; MIMIC database; PPG signal; blood pressure estimation; body area networks; compressed sensing approach; continuous operation; heart rate estimation; noninvasive monitoring; photodetector output; photoplethysmogram; power 20 mW to 60 mW; robust noise; ultra low power pulse oximeter sensor; Body area networks; Body sensor networks; Compressed sensing; Heart rate; Heart rate measurement; Light emitting diodes; Noise robustness; Pulse compression methods; Sensor phenomena and characterization; Wireless sensor networks; BAN; Blood pressure; Compressed sensing; Low power sensors; Pulse oximeter;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Wearable and Implantable Body Sensor Networks, 2009. BSN 2009. Sixth International Workshop on
Conference_Location :
Berkeley, CA
Print_ISBN :
978-0-7695-3644-6
Type :
conf
DOI :
10.1109/BSN.2009.32
Filename :
5226900
Link To Document :
بازگشت