DocumentCode
3208266
Title
Validation of a wrist monitor for accurate estimation of RR intervals during sleep
Author
Renevey, Ph ; Sola, Josep ; Theurillat, P. ; Bertschi, M. ; Krauss, J. ; Andries, D. ; Sartori, Claudio
Author_Institution
CSEM - Centre Suisse d´Electron. et Microtech., Neuchâtel, Switzerland
fYear
2013
fDate
3-7 July 2013
Firstpage
5493
Lastpage
5496
Abstract
While the incidence of sleep disorders is continuously increasing in western societies, there is a clear demand for technologies to asses sleep-related parameters in ambulatory scenarios. The present study introduces a novel concept of accurate sensor to measure RR intervals via the analysis of photo-plethysmographic signals recorded at the wrist. In a cohort of 26 subjects undergoing full night polysomnography, the wrist device provided RR interval estimates in agreement with RR intervals as measured from standard electrocardiographic time series. The study showed an overall agreement between both approaches of 0.05 ± 18 ms. The novel wrist sensor opens the door towards a new generation of comfortable and easy-to-use sleep monitors.
Keywords
biomedical equipment; electrocardiography; medical disorders; medical signal processing; parameter estimation; patient monitoring; photoplethysmography; sleep; time series; RR interval estimation; ambulatory scenarios; easy-to-use sleep monitors; full night polysomnography; photoplethysmographic signal recording; sleep disorders; sleep-related parameters; standard electrocardiographic time series; wrist monitor validation; wrist sensor; Biomedical monitoring; Electrocardiography; Monitoring; Skin; Sleep apnea; Time series analysis; Wrist;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location
Osaka
ISSN
1557-170X
Type
conf
DOI
10.1109/EMBC.2013.6610793
Filename
6610793
Link To Document