Artifact reduction with monopolar nerve cuff recording electrodes
Author
Thomsen, Morten ; Struij, Johannes Jan ; Sinkjaer, Thomas
Author_Institution
Aalborg Univ., Denmark
Volume
1
fYear
1996
fDate
31 Oct-3 Nov 1996
Firstpage
367
Abstract
In the present work an electrode configuration is presented that makes it possible to obtain reduction of artifacts when recording nerve signals monopolarly. Recordings were done in three different ways, tripolar, monopolar, and optimised monopolar. When comparing an optimised monopolar with a standard monopolar recording, where no artifact reduction takes place, stimulation artifacts are reduced to 16% and muscle artefacts to 14% of peak-peak amplitude
Keywords
bioelectric potentials; biological techniques; electrodes; neurophysiology; animal research; artifact reduction; biological research technique; electrode configuration; monopolar nerve cuff recording electrodes; muscle artefacts; neuroscience method; optimised monopolar recording; peak-peak amplitude; standard monopolar recording; stimulation artifacts; tripolar recording; Animals; Differential amplifiers; Electrical stimulation; Electrodes; Impedance measurement; Implants; Muscles; Petroleum; Rabbits; Time measurement;