DocumentCode :
3211687
Title :
A Method to Calibrate Beam Position Monitor at HLS 200 MeV Linac
Author :
Li, J.H. ; Sun, B.G. ; He, D.H. ; Lu, P. ; Wang, Y. ; Wang, J.H. ; Jin, K. ; Cao, Y. ; Zheng, P.
Author_Institution :
NSRL, University of Science and Technology of China, Hefei, Anhui 230029, P.R. China, jhli@mail.ustc.edu.cn
fYear :
2005
fDate :
16-20 May 2005
Firstpage :
2896
Lastpage :
2898
Abstract :
In order to improve injection efficiency of HLS 200Mev LINAC, we redesign a new strip line beam position monitor system, which is consisted of a strip line structure and a signal processing system. We decide on an online calibration method based on beam to find out the geometry centre displacement and relative gain offset. Before the BPM testing bench has been prepared, we make a simulation based on the model accounted for all factors influencing signal amplitudes and get the calibrating results. At last, we analyze the nonlinearity effect on the calibration results.
Keywords :
Calibration; Electrodes; High level synthesis; Impedance; Linear particle accelerator; Monitoring; Particle beams; Signal processing; Stripline; Strips;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Particle Accelerator Conference, 2005. PAC 2005. Proceedings of the
Print_ISBN :
0-7803-8859-3
Type :
conf
DOI :
10.1109/PAC.2005.1591306
Filename :
1591306
Link To Document :
بازگشت