DocumentCode
3216594
Title
Evidence of nonlinear response of hydrocephalus shunt valves to pulsating flow
Author
Nwosuh, I.A. ; Roberts ; Kosiniski ; Ciampa, A. ; Thomas, Ginu A.
Author_Institution
Dept. of Phys., New Jersey Inst. of Technol., Newark, NJ, USA
fYear
2005
fDate
2-3 April 2005
Firstpage
20
Lastpage
21
Abstract
We found that shunt valves used in regulating the cerebrospinal fluid flow (CSF) of the hydrocephalus patients alter the characteristics of the input waveform under AC flow; we tested Codman Hakim and Medtronic Strata programmable valves. We found that these two valves alter the characteristics of the fluid waveform; there is a change in the amplitude, phase and form. We are still investigating how the response of the shunt valves to AC signals similar to those present in the brain could contribute or lead to some of the complications observed in patients.
Keywords
brain; diseases; haemodynamics; prosthetics; pulsatile flow; valves; AC flow; Codman Hakim valve; Medtronic Strata programmable valve; brain; cerebrospinal fluid flow regulation; fluid waveform characteristics; hydrocephalus shunt valves; nonlinear response; pulsating flow; Absorption; Atmospheric measurements; Biomedical engineering; Fluid flow; Frequency estimation; Physics; Pressure measurement; Reservoirs; Testing; Valves;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioengineering Conference, 2005. Proceedings of the IEEE 31st Annual Northeast
Print_ISBN
0-7803-9105-5
Electronic_ISBN
0-7803-9106-3
Type
conf
DOI
10.1109/NEBC.2005.1431906
Filename
1431906
Link To Document