DocumentCode
3219363
Title
X-ray tubes incorporating carbon nanotube cathodes
Author
Espinosa, R.J. ; McKenzie, C. ; Munson, M. ; Snyder, S. ; Blake, D. ; Delzeit, L. ; Sarrazin, P.
fYear
2004
fDate
27-29 April 2004
Firstpage
253
Lastpage
254
Abstract
Oxford Instruments is currently producing tubes with CNT (carbon nanotube) cathodes for compact X-ray sources such as the 30 kV, 12.2 in3 (16.4 cm3) source. This paper provides information on the emission stability, leakage current and reproducibility of the X-ray tubes, and the effects cathode performance has on the tubes. We will also include data from an analyzer that was constructed to measure the angular divergence of the emission from these particular CNT cathodes. These data indicate that the half-angle dispersion of electrons emitted into a uniform electric field is approximately 3°.
Keywords
X-ray tubes; carbon nanotubes; cathodes; electron emission; 30 kV; C; CNT cathodes; X-ray tubes; carbon nanotube cathodes; cathode performance effects; compact X-ray source reproducibility; electron half-angle dispersion; emission angular divergence; emission stability; leakage current; Carbon nanotubes; Cathodes; Current density; Electron beams; Electron optics; Instruments; Leakage current; NASA; Stability; Structural beams;