DocumentCode
3225387
Title
Blood flow and pressure in a brachial artery under cuff pressure
Author
Kim, Ho Jong ; Lee, Hyeon Chul ; Kim, Chul Han ; Kim, Yun Jin ; Jeon, Gye Rok
Author_Institution
Dept. of CAD, Doowon Tech. Coll., Kyeonggi, South Korea
Volume
3
fYear
2004
fDate
2-6 Nov. 2004
Firstpage
2973
Abstract
In the present work, we investigate the artery pressure across a brachial artery under the cuff pressure. Deformation of the arm tissue is numerically simulated with the finite element method. The artery pressure is analyzed based on hemodynamics of blood circulation. From the analysis, it is found that the cuff pressure deforms the arm tissue significantly, but not the artery wall. Deformation of the artery wall is not significantly affected by the artery pressure. The artery pressure is found not to decrease significantly across a part of the arm pressurized by the cuff.
Keywords
biomechanics; blood flow measurement; blood pressure measurement; blood vessels; deformation; finite element analysis; arm tissue; artery pressure; artery wall; blood circulation; blood flow; blood pressure; brachial artery; cuff pressure; deformation; finite element method; hemodynamics; numerical simulation; Arterial blood circulation; Arteries; Blood flow; Blood pressure; Brachytherapy; Deformable models; Fluctuations; Hemodynamics; Numerical simulation; Veins;
fLanguage
English
Publisher
ieee
Conference_Titel
Industrial Electronics Society, 2004. IECON 2004. 30th Annual Conference of IEEE
Print_ISBN
0-7803-8730-9
Type
conf
DOI
10.1109/IECON.2004.1432284
Filename
1432284
Link To Document